260 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
[:SENSe][:FIMPedance]:CONTact1:VERify[:STATe]
Syntax [:SENSe][:FIMPedance]:CONTact1:VERify[:STATE] {ON|OFF|1|0}
[:SENSe][:FIMPedance]:CONTact1:VERify[:STATe]?
Description Determines whether to use the contact check function.
Parameters
Query response {1|0}<newline><^END>
Related commands [:SENSe][:FIMPedance]:CONTact1:VERify:THReshold1 on page 261
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold2 on page 262
Equivalent key
sequence
[Meas Setup] - CONT CHK1- ON|OFF
Contact check function is not available at 1 MHz measurement frequency.
Description
ON or 1 Enables the contact check function.
OFF or 0 (initial value) Disables the contact check function.