270 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
:STATus:OPERation:ENABle
Syntax :STATus:OPERation:ENABle <numeric>
:STATus:OPERation:ENABle?
Description Sets the value of the Operation Status Enable register.
Parameters
Query response {numeric}<newline><^END>
Related commands *SRE on page 162
:STATus:PRESet on page 271
Equivalent key
sequence
No equivalent key is available on the front panel.
:STATus:OPERation[:EVENt]?
Syntax :STATus:OPERation[:EVENt]?
Description Reads out the value of the Operation Status Event register. (Query only)
Query response {numeric}<newline><^END>
Related commands *CLS on page 157
Equivalent key
sequence
No equivalent key is available on the front panel.
<numeric>
Description The value of the Enable register.
Range 0 to 32767
Initial value 0
Resolution 1