266 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
[:SENSe][:FIMPedance]:RANGe[:UPPer]
Syntax [:SENSe][:FIMPedance]:RANGe[:UPPer]
{1p|2.2p|4.7p|10p|22p|47p|100p|220p|470p|1n|2.2n|4.7n|10n|22n|47n|100n|220
n|470n|1µ|2.2µ|4.7µ|10µ|22µ|47µ|100µ|220µ|470µ|1m}
[:SENSe][:FIMPedance]:RANGe[:UPPer]?
Description Sets the measurement range.
When you set the measurement range with this command, the measurement
range mode is automatically set to the hold range (specified to OFF with the
[:SENSe][:FIMPedance]:RANGe:AUTO command).
Parameters
If one of the settable values is not specified for the parameter, a suitable
measurement range whose recommended range includes the specified
parameter (for example, 10E-9 if the specified parameter is 5E-9) is set.
MAX or MIN can be used to specify the parameters.
Query response {1pF|2.2pF|4.7pF|10pF|22pF|47pF|100pF|220pF|470pF|1nF|2.2nF|4.7nF|10nF|2
2nF|47nF|100nF|220nF|470nF|1µF|2.2µF|4.7µF|10µF|22uF|47µF|100µF|220µF|4
70µF|1mF} <newline><^END>
If the measurement range mode is Auto, the measurement range used in the
immediately preceding measurement is read out as the query response.
Related commands [:SENSe][:FIMPedance]:RANGe:AUTO on page 265
Equivalent key
sequence
[Meas Setup] - RANGE - AUTO|HOLD
The measurement range varies depending upon the frequency setting.
<numeric>
Description Measurement Range
Range
120 Hz: 10 nF - 1 mF
1 kHz: 100 pF - 100 µF
1 MHz: 1 pF-1 nF
Initial Value 100 µF