264 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
Related commands [:SENSe][:FIMPedance]:CREJect[:STATe] on page 264
[:SENSe][:FIMPedance]:RANGe[:UPPer] on page 266
[:SENSe][:FIMPedance]:RANGe:AUTO on page 265
Equivalent key
sequence
[Meas Setup] - LOW C REJ - INCR+|DECR-
[:SENSe][:FIMPedance]:CREJect[:STATe]
Syntax [:SENSe][:FIMPedance]:CREJect[:STATe] {ON|OFF|1|0}
[:SENSe][:FIMPedance]:CREJect[:STATe]?
Description Enables/disables the Low C reject function.
When you enable the Low C reject function, if the measured value of the
primary parameter (Cp or Cs) is too small (equal to or less than the boundary
value specified with the [:SENSe][:FIMPedance]:CREJect:LIMit
command), Low C is detected. For information on the screen display,
GPIB/USB/LAN output, and so on when Low C is detected, refer to Appendix
D, “Operations when Overload, No Contact, or Low C is Detected,”.
Parameters
Query response {1|0}<newline><^END>
Related commands [:SENSe][:FIMPedance]:CREJect:LIMit on page 263
Equivalent key
sequence
[Meas Setup] - LOW C REJ - ON|OFF
Description
ON or 1 Enables the Low C reject function.
OFF or 0 (initial
value)
Disables the Low C reject function.