288 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
:TEST:HANDler:BIN
Syntax :TEST:HANDler:BIN <numeric>
Description This command sets the Handler Bin No for test purpose and can be used to
check/troubleshoot the Test pin signal of Handler interface. (No query)
Parameters
Related Commands :TEST:HANDler:COMP on page 289
:TEST:HANDler:KEYLock? on page 289
:TEST:HANDler:MODE on page 290
Equivalent key
sequence
No equivalent key is available on the front panel.
When :TEST:HANDler:MODE is set to ON, the interface pin signal value
can be controlled/read by: TEST:HAND:xxxx commands. After completing
the test, :TEST:HANDler:MODE should set to OFF so that handler interface
pin signal is generated according to the actual measurement result.
Numeric Value Description
0 Out of Bins
1-9 BIN 1 to 9
10 AUX BIN
11 All Off