290 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
:TEST:HANDler:MODE
Syntax :TEST:HANDler:MODE {ON|OFF|1|0}
:TEST:HANDler:MODE?
Description This command sets the Handler Comparator Function value for test purpose
and can be used to check/troubleshoot the Test pin signal of Handler
interface.
Parameters
Query response {1|0}<newline><^END>
Related commands :TEST:HANDler:BIN on page 288
:TEST:HANDler:COMP on page 289
:TEST:HANDler:KEYLock? on page 289
Equivalent key
sequence
No equivalent key is available on the front panel.
When :TEST:HANDler:MODE is set to ON, the interface pin signal value
can be controlled/read by: TEST:HAND:xxxx commands. After completing
the test, :TEST:HANDler:MODE should set to OFF so that handler interface
pin signal is generated according to the actual measurement result.
Description
ON or 1 Sets Handler mode to ON.
OFF or 0 (initial value) Sets Handler mode to OFF.