EasyManua.ls Logo

Keysight E4981A - Calculate1:Comparator:auxbin

Keysight E4981A
389 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
172 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
:CALCulate1:COMParator:AUXBin
Syntax :CALCulate1:COMParator:AUXBin {ON|OFF|1|0}
:CALCulate1:COMParator:AUXBin?
Description Turns ON/OFF the AUX_BIN function for sorting of the comparator function.
Depending on the ON/OFF state of the AUX_BIN function, the following
difference occurs in the sorting result when the measurement result of the
secondary parameter exceeds the limit range.
ON: Sorted into AUX_BIN if the measurement result of the primary
parameter is within the limit range. Otherwise, sorted into
OUT_OF_BINS.
OFF: Always sorted into OUT_OF_BINS.
Parameters
Query response {1|0}<newline><^END>
Equivalent key
sequence
[Meas Setup] - LIMIT TABLE - AUX
Description
ON or 1 Turns ON the AUX_BIN function.
OFF or 0 (initial value) Turns OFF the AUX_BIN function.

Table of Contents

Other manuals for Keysight E4981A

Related product manuals