172 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
:CALCulate1:COMParator:AUXBin
Syntax :CALCulate1:COMParator:AUXBin {ON|OFF|1|0}
:CALCulate1:COMParator:AUXBin?
Description Turns ON/OFF the AUX_BIN function for sorting of the comparator function.
Depending on the ON/OFF state of the AUX_BIN function, the following
difference occurs in the sorting result when the measurement result of the
secondary parameter exceeds the limit range.
ON: Sorted into AUX_BIN if the measurement result of the primary
parameter is within the limit range. Otherwise, sorted into
OUT_OF_BINS.
OFF: Always sorted into OUT_OF_BINS.
Parameters
Query response {1|0}<newline><^END>
Equivalent key
sequence
[Meas Setup] - LIMIT TABLE - AUX
Description
ON or 1 Turns ON the AUX_BIN function.
OFF or 0 (initial value) Turns OFF the AUX_BIN function.