Loading...
226 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
:HCOPy:SDUMp:DATA?
Syntax :HCOPy:SDUMp:DATA?
Description Outputs screen image display in GIF format. (Query Only)
Equivalent key No equivalent key is available on the front panel.
Example of Use
Dim Nop As Long
Dim GifData(10000) As Byte
Dim paramsArray(2) As Long
Dim i As Integer
Nop = UBound(GifData) - LBound(GifData) + 1
paramsArray(0) = VarPtr(Nop)
paramsArray(1) = VarPtr(GifData(0))
Call viVPrintf(Agte4981a, ":HCOPY:SDUMP:DATA?" + vbLf, 0)
Call viVScanf(Agte4981a, "%#b", paramsArray(0))
Open "C:\TEST.gif" For Binary As #1
For i = 0 To Nop - 1
Put #1, , GifData(i)
Next i
Close
Pressing [Save/Recall] - SAVE DISPLAY key saves the current screen image
into the USB memory.