252 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
[:SENSe][:FIMPedance]:APERture[:MODE]
Syntax [:SENSe][:FIMPedance]:APERture[:MODE] {SHORt|MEDium|LONG}
[:SENSe][:FIMPedance]:APERture[:MODE]?
Description Selects the measurement time (integral time) mode from SHORt, MEDium or
LONG. This command is provided to support 4268A/4288A commands.
When Short is selected, the [:SENSe][:FIMPedance]:APERture:TIME is set
at 1, MED is set at 4, LONG is set at 8.
The Query command returns SHORt when
[:SENSe][:FIMPedance]:APERture:TIME command is set to 1 or 2, returns
MED when APER:TIME is set at 4, and returns LONG when APER:TIME is set at
6 or 8.
For information on the specific measurement time of each mode, see
“Specification and Supplemental Performance Characteristics” in the
Operation Manual.
Parameters
Query response {SHOR|MED|LONG}<newline><^END>
Equivalent key
sequence
No equivalent key can be used on front panel.
Description
SHORt (initial value) Specifies the short mode (Aperture Time=1 or 2).
MEDium Specifies the medium mode (Aperture Time=4).
LONG Specifies the long mode (Aperture Time=6 or 8)