244 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
[:SENSe]:CORRection:MULTiple:CKIT:STAN3[:STATe]
Syntax [:SENSe]:CORRection:MULTiple:CKIT:STAN3[:STATe] {ON|OFF|1|0}
[:SENSe]:CORRection:MULTiple:CKIT:STAN3[:STATe]?
Description Determines whether to define the standard values for LOAD correction for each
channel individually when using the multi-correction function (set to ON with
the [:SENSe]:CORRection:MULTiple[:STATe] command).
Parameters
Query response {1|0}<newline><^END>
Related commands [:SENSe]:CORRection:MULTiple[:STATe] on page 245
[:SENSe]:CORRection:CKIT:STAN3[:DATA] on page 235
Equivalent key
sequence
[Meas Setup] - CORRECTION - LOAD REF - MULTI|SINGLE
Description
ON or 1 Enables channel-by-channel value definition.
OFF or 0
(initial value)
Disables channel-by-channel value definition (defines the same
values for all channels).