180 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
:CALCulate1:COMParator:COUNt[:STATe]
Syntax :CALCulate1:COMParator:COUNt[:STATe] {ON|OFF|1|0}
:CALCulate1:COMParator:COUNt[:STATe]?
Description Turns ON/OFF the BIN counter function of the comparator function.
If you turn on this function, the number of DUTs sorted into each BIN based on
the comparator sorting result is counted. The maximum count is 999999. If this
is exceeded, the count value will not be updated but remain at 999999.
Parameters
Query response {1|0}<newline><^END>
Related commands :CALCulate1:COMParator:COUNt:CLEar on page 175
:CALCulate1:COMParator:COUNt:DATA? on page 176
:CALCulate1:COMParator:COUNt:OVLD? on page 179
:CALCulate1:COMParator:COUNt:MULTiple:DATA? on page 177
:CALCulate1:COMParator:COUNt:MULTiple:OVLD? on page 179
Equivalent key
sequence
[Display Format] - BIN COUNT - COUNT - COUNT ON|COUNT OFF
Description
ON or 1 Turns ON the BIN counter function.
OFF or 0 (initial value) Turns OFF the BIN counter function.