EasyManua.ls Logo

Keysight E4981A - Calculate2:Math:expression:catalog; Calculate2:Math:expression:name

Keysight E4981A
389 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
192 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
:CALCulate2:MATH:EXPRession:CATalog?
Syntax :CALCulate2:MATH:EXPRession:CATalog?
Description Reads out available parameters used when executing the command to specify
the expression of the secondary parameter in the deviation measurement mode
(the :CALCulate1:MATH:EXPRession:NAME command). The query response
is always DEV,PCNT. (Query only)
Query response DEV,PCNT<newline><^END> (fixed)
Related commands :CALCulate2:MATH:EXPRession:NAME on page 192
Equivalent key
sequence
No equivalent key is available on the front panel.
:CALCulate2:MATH:EXPRession:NAME
Syntax :CALCulate2:MATH:EXPRession:NAME {DEV|PCNT}
:CALCulate2:MATH:EXPRession:NAME?
Description Specifies the expression of the secondary parameter used when displaying the
measurement result in the deviation measurement mode.
Parameters
Query response {DEV|PCNT}<newline><^END>
Related commands :CALCulate2:MATH:STATe on page 193
:CALCulate1:MATH:EXPRession:NAME on page 189
:DATA[:DATA] on page 208
Equivalent key
sequence
[Meas Setup] - DEV B - delta ABS|delta %
Description
DEV
(initial value)
Displays the result as the difference between the measurement
value and the reference value
1
(measurement value - reference
value).
PCNT Displays the difference between the measurement value and the
reference value
*1
as a percentage
2
of the reference value.
1. Use the :DATA[:DATA] command to set the reference value.
2. (Measurement value - reference value) / reference value×100

Table of Contents

Other manuals for Keysight E4981A

Related product manuals