182 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
:CALCulate1:COMParator:PRIMary:BIN{1-9}[:LIMIT]
Syntax :CALCulate1:COMParator:PRIMary:BIN{1|2|3|4|5|6|7|8|9}[:LIMIT] <numeric
1>,<numeric 2>
:CALCulate1:COMParator:PRIMary:BIN{1|2|3|4|5|6|7|8|9}[:LIMIT]?
Description Sets the low and high limit ranges of BIN1 to BIN9 as the primary parameter
used in the comparator function, depending on the designation method set
using the :CALCulate1:COMParator:MODE command.
This command only sets the limit range. To activate it, use the
:CALCulate1:COMParator:PRIMary:BIN{1-9}:STATe command to turn on
the BIN for which the limit range has been set.
Parameters
If the specified parameter is out of the allowable setup range, the minimum
value (if the lower limit of the range is not reached) or the maximum value (if
the upper limit of the range is exceeded) is set.
Query response {numeric 1},{numeric 2}<newline><^END>
Related commands :CALCulate1:COMParator:MODE on page 181
:CALCulate1:COMParator:PRIMary:BIN{1-9}:STATe on page 183
:CALCulate1:COMParator:SECondary:LIMit on page 185
:CALCulate1:COMParator[:STATe] on page 187
Equivalent key
sequence
[Meas Setup] - LIMIT TABLE - BIN{1|2|3|4|5|6|7|8|9}- LOW, HIGH
<numeric 1> <numeric 2>
Description The lower limit value of the limit
range.
The upper limit value of the
limit range.
Range -999.999 to 999.999 -999.999 to 999.999
Initial value 0 0
Unit F (farad) or % (percent) F (farad) or % (percent)
The unit of the parameter may change depending on the limit range
designation method.