250 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
[:SENSe]:DETector:DELay1
Syntax [:SENSe]:DETector:DELay1 <numeric>
[:SENSe]:DETector:DELay1?
Description Sets/Gets the waiting time for analog measurement for 120 Hz Frequency
measurement.
Parameters
Query response {numeric}<newline><^END>
Related commands [:SENSe]:DETector:DELay2 on page 250
[:SENSe]:DETector:DELay3 on page 251
Equivalent key
sequence
No equivalent key can be used on front panel.
[:SENSe]:DETector:DELay2
Syntax [:SENSe]:DETector:DELay2 <numeric>
[:SENSe]:DETector:DELay2?
Description Sets/Gets the waiting time for analog measurement for 1 kHz Frequency
measurement.
Parameters
Query response {numeric}<newline><^END>
<numeric 1>
Description Waiting time for 120 Hz measurement frequency
Range 0 to 100m
Initial value 1.67m
Resolution 1u
<numeric 1>
Description Waiting time for 1 kHz measurement frequency
Range 0 to 100m
Initial value 1m
Resolution 1u