EasyManua.ls Logo

Keysight E4981A - Source:frequency[:Cw]; Source:voltage:alc[:State]

Keysight E4981A
389 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 267
Command Reference
E4981A SCPI Commands
:SOURce:FREQuency[:CW]
Syntax :SOURce:FREQuency[:CW] <numeric> [Hz|kHz|MHz]
:SOURce:FREQuency[:CW]?
Description Sets the measurement frequency.
Depending on this setting, the available measurement range varies (set with
the [:SENSe][:FIMPedance]:RANGe[:UPPer] command). If changing the
measurement frequency results in a conflict with the measurement range
setting, an acceptable range is automatically selected.
Parameters
Query response {120|1E3|1E6}<newline><^END>
Related commands [:SENSe][:FIMPedance]:RANGe[:UPPer] on page 266
Equivalent key
sequence
[Meas Setup] - FREQ - 120 Hz|1 kHz|1 MHz
:SOURce:VOLTage:ALC[:STATe]
Syntax :SOURce:VOLTage:ALC[:STATe] {ON|OFF|1|0}
:SOURce:VOLTage:ALC[:STATe]?
Description Determines whether to use the signal level compensation function (LVL
COMP).
Parameters
Query Response {1|0}<newline><^END>
Equivalent key
sequence
[Meas Setup] - LVL COMP- ON|OFF
Description
Range Option 001 = 120 Hz, 1kHz and 1MHz
Option 002 = 120 Hz and 1 kHz
Initial value 1 kHz
Description
ON or 1 Enables the signal level compensation function.
OFF or 0 (Initial value) Disables the signal level compensation function.

Table of Contents

Other manuals for Keysight E4981A

Related product manuals