Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 301
Command Reference
E4981A SCPI Commands
:TEST:SCANner:MODE
Syntax :TEST:SCANner:MODE {ON|OFF|1|0}
:TEST:SCANner:MODE?
Description This command sets the Scanner Test Mode ON/OFF for test purpose and can
be used to check/troubleshoot the Test pin signal of Scanner interface.
Parameters
Query response {1|0}<newline><^END>
Related commands :TEST:SCANner:CH? on page 298
:TEST:SCANner:EOM on page 299
:TEST:SCANner:INDex on page 300
:TEST:SCANner:TRIGger? on page 302
:TEST:SCANner:VALid? on page 302
Equivalent key
sequence
No equivalent key is available on the front panel.
When :TEST:SCANner:MODE is set to ON, the interface pin signal value
can be controlled/read by: TEST:SCAN:xxxx commands. After completing
the test, :TEST:SCANner:MODE should set to OFF so that scanner interface
pin signal is generated according to the actual measurement result.
Description
ON Sets Scanner mode to ON
OFF Sets Scanner mode to OFF.