Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 271
Command Reference
E4981A SCPI Commands
:STATus:OPERation:UPDate
Syntax :STATus:OPERation:UPDate {ON|OFF|1|0}
:STATus:OPERation:UPDate?
Description Enables/Disables update of the Operation Status Event register. The disabling
of status register can shorten the EOM time.
Parameters
Query response {1|0}<newline><^END>
Related commands :STATus:OPERation:ENABle on page 270
Equivalent key
sequence
No equivalent key is available on the front panel.
:STATus:PRESet
Syntax :STATus:PRESet
Description Initializes the Operation Status register and the Questionable Status register.
(No query)
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
ON or 1 Enables the update of the Operation Status Event register.
OFF or 0 (initial value) Disables the update of the Operation Status Event register.