262 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold2
Syntax [:SENSe][:FIMPedance]:CONTact1:VERify:THReshold2 <numeric>
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold2?
Description Sets/Gets the Threshold2 value for Contact check. Threshold2 is a parameter
which is proportional contact resistance of Lp or Lc. The recommended setting
value is described in the Operation Manual.
Parameters
Query response <numeric><newline><^END>
Related commands [:SENSe][:FIMPedance]:CONTact1:VERify[:STATe] on page 260
[:SENSe][:FIMPedance]:CONTact1:VERify:THReshold1 on page 261
Equivalent key
sequence
[Meas Setup] - CONT CHK- CC1 TH2
The
contact failure at Lp can not be detected at the range of 220 uF to 1 mF
at 120 Hz and 22 uF to 100 uF at 1 kHz. At the same range, the threshold1
is used instead of this for Lc. (Therefore, the threshold2 is not used at the
range).
<numeric>
Description Threshold2 value for contact 1
Range 0 to 1
Initial value 1
Resolution 0.01