Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 247
Command Reference
E4981A SCPI Commands
[:SENSe]:CORRection:OFFSet[:STATe]
Syntax [:SENSe]:CORRection:OFFSet[:STATe] {ON|OFF|1|0}
[:SENSe]:CORRection:OFFSet[:STATe]?
Description Turns ON/OFF the offset correction function.
Parameters
Query response {1|0}<newline><^END>
Related commands [:SENSe]:CORRection:OFFSet:DATA on page 246
:CALCulate1:FORMat on page 188
:CALCulate2:FORMat on page 191
Equivalent key
sequence
[Meas Setup] - CORRECTION - OFFSET -ON|OFF
Changing any of the measurement parameter will automatically turn off
this function. To avoid this, you have to write a program so that this
command is executed after the execution of the measurement parameter
setup command (:CALCulate1:FORMat and :CALCulate2:FORMat).
Description
ON or 1 Turns ON the offset correction function.
OFF or 0 (initial value) Turns OFF the offset correction function.