EasyManuals Logo

Keysight E4981A Programming Manual

Keysight E4981A
389 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #247 background imageLoading...
Page #247 background image
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 247
Command Reference
E4981A SCPI Commands
[:SENSe]:CORRection:OFFSet[:STATe]
Syntax [:SENSe]:CORRection:OFFSet[:STATe] {ON|OFF|1|0}
[:SENSe]:CORRection:OFFSet[:STATe]?
Description Turns ON/OFF the offset correction function.
Parameters
Query response {1|0}<newline><^END>
Related commands [:SENSe]:CORRection:OFFSet:DATA on page 246
:CALCulate1:FORMat on page 188
:CALCulate2:FORMat on page 191
Equivalent key
sequence
[Meas Setup] - CORRECTION - OFFSET -ON|OFF
Changing any of the measurement parameter will automatically turn off
this function. To avoid this, you have to write a program so that this
command is executed after the execution of the measurement parameter
setup command (:CALCulate1:FORMat and :CALCulate2:FORMat).
Description
ON or 1 Turns ON the offset correction function.
OFF or 0 (initial value) Turns OFF the offset correction function.

Table of Contents

Other manuals for Keysight E4981A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight E4981A and is the answer not in the manual?

Keysight E4981A Specifications

General IconGeneral
BrandKeysight
ModelE4981A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals