EasyManua.ls Logo

Keysight E4981A - System:scanner:trigger:voltage

Keysight E4981A
389 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 285
Command Reference
E4981A SCPI Commands
:SYSTem:SCANner:TRIGger:VOLTage
Syntax :SYSTem:SCANner:TRIGger:VOLTage <numeric>
:SYSTem:SCANner:TRIGger:VOLTage?
Description Sets the scanner trigger input voltage.
Parameters
Query response {numeric}<newline><^END>
Equivalent key No equivalent key is available on the front panel.
<numeric>
Description The value of the scanner trigger input voltage
Range 5 to 15
Initial (factory set)
value
15
Unit V (voltage)
Resolution 100m
The value of this command is not changed with *RST and :SYSTem:PRESet
and is only changed with Factory Default reset (available only through
Front Panel).
The product may get damaged if incorrect voltage is set to Scanner
Trigger. Take caution while changing the Scanner Trigger voltage.