302 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
:TEST:SCANner:TRIGger?
Syntax :TEST:SCANner:TRIGger?
Description This command gets the Scanner Trigger signal for test purpose and can be
used to check/troubleshoot the Test pin signal of Scanner interface. (Query
Only)
Query response {1|0}<newline><^END>
Related commands :TEST:SCANner:CH? on page 298
:TEST:SCANner:EOM on page 299
:TEST:SCANner:INDex on page 300
:TEST:SCANner:MODE on page 301
:TEST:SCANner:VALid? on page 302
Equivalent key
sequence
No equivalent key is available on the front panel.
:TEST:SCANner:VALid?
Syntax :TEST:SCANner:VALid?
Description Gets scanner/CH_VALID signal.(Query only)
Query response {1|0}<newline><^END>
Related commands :TEST:SCANner:CH? on page 298
:TEST:SCANner:EOM on page 299
:TEST:SCANner:INDex on page 300
:TEST:SCANner:MODE on page 301
:TEST:SCANner:TRIGger? on page 302
Equivalent key
sequence
No equivalent key is available on the front panel.
When :TEST:SCANner:MODE is set to ON, the interface pin signal value
can be controlled/read by: TEST:SCAN:xxxx commands. After completing
the test, :TEST:SCANner:MODE should set to OFF so that scanner interface
pin signal is generated according to the actual measurement result.