Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 209
Command Reference
E4981A SCPI Commands
If the specified parameter is out of the allowable setup range, the minimum
value (if the lower limit of the range is not reached) or the maximum value (if
the upper limit of the range is exceeded) is set.
Query response
When specifying BUF1 or BUF2 as parameter
{numeric 1},{numeric 2},{numeric 3},…,{numeric (N×3)}<newline><^END>
The N sets of measurement data (a data set consists of the measurement
status, measured value, and comparator sorting result) fed into the data buffer
are read out in the order of measurement.
Where N is the number of measurement points specified with the
:DATA:POINts:BUF1or :DATA:POINts:BUF2 command and n is an integer
between 1and N:
<numeric>
Description Reference value used in the deviation measurement mode
Range -99.9999E9 to 99.9999E9 (for REF1)
-99.9999E9 to 99.9999E9 (for REF2)
Initial value 0
Unit Depends on the type of parameter.
{numeric (n–1)×3+1}
The measurement status in the n-th measurement (an integer
between 0 and 3 as shown below).
0: No error
1: Detection of measurement impossibility (overload)
2: Detection of Low C reject or No Contact
{numeric (n–1)×3+2}
The measured value of the primary or secondary parameter
1
in
the n-th measurement. If overload is detected, 9.9E37 is
outputted.