210 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Command Reference
E4981A SCPI Commands
10-
When specifying BUF3 as parameter (Comparator: OFF)
{numeric 1},{numeric 2},{numeric 3},…,{numeric (N×3)}<newline><^END>
The N sets of measurement data (a data set consists of the measurement
status, measured value, and comparator sorting result) fed into the data buffer
are read out in the order of measurement.
Where N is the number of measurement points specified with the
:DATA:POINts:BUF3 command and n is an integer between 1 and N:
{numeric (n–1)×3+3}
The comparator sorting result in the n-th measurement (an
integer between 0 and 11 as shown below). The output (output
value is 11) is produced even if the comparator is off.
0: Sorted into OUT_OF_BINS.
1: Sorted into BIN1.
2: Sorted into BIN2.
3: Sorted into BIN3.
4: Sorted into BIN4.
5: Sorted into BIN5.
6: Sorted into BIN6.
7: Sorted into BIN7.
8: Sorted into BIN8.
9: Sorted into BIN9.
10: AUX_BIN
11: OVLD (Overload) or NC (No contact) [sorting not
possible]
1. You need to select primary parameter or secondary parameter by using the
:DATA:FEED:BUF1 or :DATA:POINts:BUF2 command in advance.
{numeric (n–1)×3+1}
The measurement status in the n-th measurement (an integer
between 0 and 2).
{numeric (n–1)×3+2}
The measured value of the primary parameter in the n-th
measurement. If overload is detected, 9.9E37 is outputted.
{numeric (n–1)×3+3}
The measured value of the secondary parameter in the n-th
measurement. If overload is detected, 9.9E37 is outputted.