Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter 221
Command Reference
E4981A SCPI Commands
The transfer format of data read out with this command conforms to the setup
made with the :FORMat[:DATA] command. (Query only)
Query response {numeric 1},{numeric 2},{numeric 3},{numeric 4}<newline><^END>
Related commands :READ? on page 230
*TRG on page 163
:FORMat[:DATA] on page 225
:CALCulate1:COMParator[:STATe] on page 187
Equivalent key
sequence
No equivalent key is available on the front panel.
If the comparator function is disabled (OFF has been specified with the
:CALCulate1:COMParator[:STATe] command), only three data items,
{numeric1}, {numeric2}, and {numeric3}, are read out.
{numeric 1}, {numeric 2}, {numeric 3}, and {numeric 4} are as follows:
{numeric 1}: Measurement status (one of the integers between 0 and 3
listed below)
0: No error
1: Detection of overload (OVLD)
2: Detection of Low C or No Contact (See
:FORMat:STATus:EXTension on page 224)
{numeric 2}: Measured value of the primary parameter
{numeric 3}: Measured value of the secondary parameter
{numeric 4}: Comparator sorting result (an integer between 0 and 11
listed below)
0: Sorted into OUT_OF_BINS.
1: Sorted into BIN1.
2: Sorted into BIN2.
3: Sorted into BIN3.
4: Sorted into BIN4.
5: Sorted into BIN5.
6: Sorted into BIN6.
7: Sorted into BIN7.
8: Sorted into BIN8.
9: Sorted into BIN9.
10: Sorted into AUX_BIN.
11: OVLD (Overload) or NC (No contact) [sorting not possible]