374 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Operations when Overload, No Contact, or Low C is Detected
Operations when Overload/No Contact/Low C is Detected
-
Table D-5 At-a-glance table of operations when Overload/No contact/Low C reject is detected
Display output GPIB/LAN/USB output
Handler output
(handler signal
that becomes
active)
Mea-
sured
value
Voltage/
current
monitor
value
Compara-
tor sort-
ing result
Measure-
ment sta-
tus
Mea-
sured
value
Compar-
ator sort-
ing result
Overload OVLD --- --- 1 9.9E37 11 /OVLD
No contact N.C. --- --- 2 9.9E37 11 /LOWC_OR_NC
Low C Normal Normal
LOWC
1
2 Normal Normal
/LOWC_OR_NC
2
Out of display
range
3
------- --- Normal Normal Normal Normal Normal
1. When Low C is detected, LOWC is displayed in BIN No. DISPLAY page.
2. /LOWC_OR_NC becomes active together with the signal that corresponds to the result of normal sorting
judgment (judgment result when no error occurred).
3. When an out-of-display-range state is detected, measurement is performed normally but the measured val-
ues are not displayed.