70 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Reading Out Measured Result
Reading Out Measured Result
6-
Figure 6-9 Structure of data buffer
Procedure for use
Step 1. Use the :DATA:POINts:BUF1 command on page 202 to set up the number of
measurements you want to feed into the data buffer.
Step 2. When you use buffer 1 or buffer 2, use the :DATA:FEED:BUF1 command on
page 195 to select whether the parameter is fed from the primary or
secondary parameter.
Step 3. Use the :DATA:FEED:CONTrol[:STATe] command on page 200 to make the
setup feed the measured result into the data buffer.
Step 4. Execute the measurement for the number of times specified in Step 1.
Step 5. Use the :DATA[:DATA] command on page 208 to read out the data
temporarily stored in the data buffer.
Step 6. Repeat Steps 4 to 5.
Sample program
See Example 9-12 to view an example of reading out measured result using
data buffer.