74 Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Sorting Based on Measured Result (Comparator Function)
Setting Up Comparator Function
7-
Selecting limit range designation method
You can select the designation method of the primary parameter limit ranges
(BIN1 to BIN9) from the following three modes:
To select the designation method, use the following command.
:CALCulate1:COMParator:MODE on page 181
To set up the reference value for the absolute tolerance mode or percent
tolerance mode, use the following command.
:CALCulate1:COMParator:PRIMary:NOMinal on page 184
Turning ON/OFF limit range
To turn ON/OFF the primary parameter limit range (BIN1 to BIN9), use the
following command.
:CALCulate1:COMParator:PRIMary:BIN{1-9}:STATe on page 183
Only BINs set to ON with the above command become the targets of sorting
judgment. In other words, even if the measured result falls within the limit
range of a BIN set to OFF, it will not be sorted to that BIN.
To turn ON/OFF the secondary parameter limit range, use the following
command.
:CALCulate1:COMParator:SECondary:STATe on page 186
If you specify OFF with the above command, no sorting judgment is made for
the measured result of the secondary parameter. The comparator sorting
judgment result is determined only on the basis of the measured primary
parameter result.
Mode Description
Absolute mode Designation using the absolute value
Absolute tolerance mode Designation using the
relative value
(deviation from the
reference value)
Designation using the absolute
value
1
1. Boundary value - reference value
Percent tolerance mode Designation using the percentage
of the reference value
2
2. ((Boundary value - reference value)/reference value)×100
Only the absolute mode can be used as the designation method of the
secondary parameter limit range.