Improving level performance
R&S
®
SMA100B
236User Manual 1178.3834.02 ─ 09
Remote command:
:SLISt[:LIST]? on page 549
:SLISt:ELEMent<ch>:MAPPing on page 551
:SLISt:SENSor:MAP on page 551
Scan
Scans the network and the USB connections for sensors connected using the VISA
communication protocol, i.e. sensors that are addressed over LAN or USBTMC.
The instrument detects sensors communicating over the USB legacy protocol automat-
ically.
Remote command:
:SLISt:SCAN[:STATe] on page 549
Clear
Removes the selected sensor from the sensor mapping list.
Remote command:
:SLISt:CLEar:LAN on page 550
:SLISt:CLEar:USB on page 550
:SLISt:CLEar[:ALL] on page 551
Add Sensor/Hide 'Add Sensor'
Shows or hides the "Add Sensor" settings.
Add Sensor settings
Configures settings to add sensors connected to the R&S SMA100B via USB or LAN.
Add LAN Sensor settings ← Add Sensor settings
Configures settings to add sensors connected to the R&S SMA100B via LAN.
"IP Address or Host Name"
Displays the host name or the IP address of a R&S NRP power sen-
sor.
If the R&S SMA100B does not detect a connected R&S NRP sensor,
you can assign the address information manually.
"Add LAN Sensor"
Adds a detected R&S NRP sensor connected in the LAN to the list of
sensors, including its device ID or name and its serial number.
Remote command:
:SLISt:SCAN:LSENsor on page 549
Add USB Sensor settings ← Add Sensor settings
Configures settings to add sensors connected to the R&S SMA100B via USB.
"Device ID or Sensor Name"
Displays the device identifier or the name of the R&S NRP power
sensor.
If the R&S SMA100B does not detect a connected R&S NRP sensor,
you can assign the ID or name manually.
Using power sensors