Network operation and remote control
R&S
®
SMA100B
404User Manual 1178.3834.02 ─ 09
The "Instrument Emulations" dialog enables you to emulate a remote control com-
mand set of another signal generator.
The remote commands required to configure the emulation settings remotely are
described in Chapter 14.17, "SYSTem subsystem", on page 712.
Language
Selects the instrument whose remote command set is emulated by the R&S SMA100B.
Remote command:
:SYSTem:LANGuage on page 729
Mode
Selects the way the instrument identification is performed.
"Automatic"
Sets the "IDN String" and the "OPT String" automatically for the
instrument selected with the parameter Language.
"User Defined"
Enables you to define the "IDN String" and the "OPT String".
Remote command:
:SYSTem:IDENtification on page 728
Set to Default
In "Mode > User Defined", resets the *IDN and *OPT strings.
Remote command:
:SYSTem:IDENtification:PRESet on page 729
IDN String
Indicates the identification string of the instrument when queried with the common
command *IDN?.
In addition to the preset values, you can define your own identification string so that
each generator uses an individual identification, like My_SigGen for instance, see
Mode .
Remote command:
*IDN? on page 451
:SYSTem:IRESponse on page 729
OPT String
Indicates the option string of the instrument as queried with common command *OPT?.
In Mode > "User Defined", you can define your own option string. A query responds
with this option string instead of the automatically created option string.
Remote command:
*OPT? on page 452
:SYSTem:ORESponse on page 729
Remote access settings