EasyManua.ls Logo

R&S SMA100B - Page 729

R&S SMA100B
822 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
SMA100B
729User Manual 1178.3834.02 ─ 09
:SYSTem:IDENtification:PRESet
Sets the *IDN and *OPT strings in user defined mode to default values.
Example:
SYST:IDEN USER
SYST:IDEN:PRES
Usage: Event
Manual operation: See "Set to Default" on page 404
:SYSTem:IRESponse
<IdnResponse>
Defines the user defined identification string for *IDN.
Note: While working in an emulation mode, the instrument's specific command set is
disabled, i.e. the SCPI command SYST:IRES is discarded.
Parameters:
<IdnResponse> string
Example:
SYST:IDEN USER
// Selects a user-defined identification
SYST:IRES "Test Device"
// Defines identification string 'test device'
*IDN?
// Response: 'test device'
Manual operation: See "IDN String" on page 404
:SYSTem:ORESponse <OResponse>
Defines the user defined response string for *OPT.
Note: While working in an emulation mode, the instrument's specific command set is
disabled, i.e. the SCPI command SYST:ORES is discarded.
Parameters:
<OResponse> string
Example:
SYST:IDEN USER
// Selects a user-defined identification
SYST:ORES "Test Option"
// Defines the OPT string 'test option'
*OPT?
// Response: 'test option'
Manual operation: See "OPT String" on page 404
:SYSTem:LANGuage <Language>
Sets the remote control command set.
SYSTem subsystem

Table of Contents

Other manuals for R&S SMA100B

Related product manuals