Remote control commands
R&S
®
SMA100B
473User Manual 1178.3834.02 ─ 09
Parameters:
<Phase> float
Range: -36000 to 36000
Increment: 0.1
*RST: 0
Example: See Chapter 14.7, "CSYNthesis subsystem", on page 469.
Manual operation: See "Delta Phase" on page 307
:CSYNthesis:PHASe:REFerence
Resets the delta phase value.
Example:
See Chapter 14.7, "CSYNthesis subsystem", on page 469.
Usage: Event
:CSYNthesis:POWer:STEP:MODE
<Mode>
:CSYNthesis:FREQuency:STEP:MODE <Mode>
Defines the type of step size to vary the frequency and level at discrete steps.
Parameters:
<Mode> DECimal | USER
DECimal
Increases or decreases the level in steps of 10.
USER
Increases or decreases the value in increments, set with the
command:
:CSYNthesis:FREQuency:STEP
:CSYNthesis:POWer:STEP[:INCRement]
*RST: DECimal
Example: See Chapter 14.7, "CSYNthesis subsystem", on page 469.
Manual operation: See "Variation Active" on page 307
:CSYNthesis:POWer:STEP[:INCRement] <Increment>
Sets the step width of the rotary knob and, in user-defined step mode, increases or
decreases the level.
Parameters:
<Increment> float
Range: 0 to 35
Increment: 0.01
*RST: 1
Example: See Chapter 14.7, "CSYNthesis subsystem", on page 469.
Manual operation: See "Variation Step" on page 307