Remote control commands
R&S
®
SMA100B
521User Manual 1178.3834.02 ─ 09
Manual operation: See "Clear Parth" on page 282
:SENSe[:POWer]:SWEep:HCOPy:FILE[:NAME]:AUTO:FILE?
Queries the file name generated with the automatic naming settings.
Note: As default the automatically generated file name is composed of: >PAth>/
<Prefix><YYYY><MM><DD><Number>.<Format>. Each component can be deacti-
vated/ activated separately to individually design the file name.
Return values:
<File> string
Example:
SENS:SWE:HCOP:DEV:LANG BMP
Selects output format *.bmp.
SENS:SWE:HCOP:FILE:AUTO:DIR 'var/nrp'
Hardcopy files are stored in directory var/nrp if automatic
naming is activated.
SENS:SWE:HCOP:FILE:AUTO:PREF 'sens1'
The file name starts with the prefix sens1. The usage of auto-
matic naming with prefix and date in the file name is preset
(…:STAT ON).
SENS:SWE:HCOP
Triggers the generation of a hardcopy of the current trace.
SENS:SWE:HCOP:FILE:AUTO:FILE?
Queries the file name
Usage:
Query only
Manual operation: See "File name - Power Analysis" on page 280
:SENSe[:POWer]:SWEep:HCOPy:FILE[:NAME]:AUTO:STATe <State>
Activates/deactivates automatic naming of the hardcopy files.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 1
Example:
SENS:SWE:HCOP:FILE:AUTO:STAT OFF
Deactivates automatic naming of the hardcopy files. The file
name and directory is now defined with command.
SENS:SWE:HCOP:FILE:NAME <psth>
Manual operation: See "File name - Power Analysis" on page 280
:SENSe[:POWer]:SWEep:HCOPy:FILE[:NAME]:AUTO[:FILE]:DAY?
Queries the day of the date part in the automatic file name.
Power sensor measurement subsystems