Remote control commands
R&S
®
SMA100B
621User Manual 1178.3834.02 ─ 09
EXT|INT,EXT
An external modulation source is used, additional to the internal
modulation source.
The external signal is input at the Ext connector.
*RST: INT
Example: See Example "Configure and generate an ILS localizer signal"
on page 595.
Manual operation: See "EXT AM" on page 153
[:SOURce<hw>][:ILS]:MBEacon:COMid:CODE
<Code>
Sets the coding of the COM/ID signal by the international short name of the airport
(e.g. MUC for the Munich airport).
The COM/ID tone is sent according to the selected code, see Chapter D, "Morse code
settings", on page 784.
If no coding is set, the COM/ID tone is sent uncoded (key down).
Parameters:
<Code> string
Example: See Example "Configure and generate an ILS marker beacons
signal" on page 596.
Manual operation: See "Code" on page 164
[:SOURce<hw>][:ILS]:MBEacon:COMid:DASH <Dash>
Sets the length of a Morse code dash.
Parameters:
<Dash> float
Range: 0.05 to 1
Increment: 1E-4
*RST: 0.3
Example: See Example "Configure and generate an ILS marker beacons
signal" on page 596.
Manual operation: See "Dash Length" on page 165
[:SOURce<hw>][:ILS]:MBEacon:COMid:DEPTh <Depth>
Sets the AM modulation depth of the COM/ID signal.
Parameters:
<Depth> float
Range: 0 to 100
Increment: 0.1
*RST: 5
SOURce subsystem