EasyManua.ls Logo

Capintec CAPTUS 4000e - Changing Detectors; Spectrum Scale & Zooming; Vertical Zoom

Capintec CAPTUS 4000e
370 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CAPINTEC, INC. CAPTUS
®
4000e
August 15 MCA 13-3
CHANGING DETECTORS
To switch detectors, simply select the desired detector from the Detector frame on the lower
right side of Figure 13-1 MCA Main Screen. The program will automatically switch the signal
input to the selected detector, and reset the stored calibration settings for that detector.
SPECTRUM SCALE & ZOOMING
The spectral display area will automatically change vertical scale range so that the entire
spectrum is displayed at all times.
The spectrum can be zoomed-in vertically or horizontally by selecting the Zoom button on
the menu bar. A pop-up window with three options is displayed:
Vertical ZoomThis selection will allow zooming of the full-count vertical scale.
Horizontal ZoomThis selection will allow zooming of the horizontal scale.
Auto RangeThis selection will restore the spectrum the original automatic settings.
Vertical Zoom
The vertical scale can be expanded to a (minimum) full-scale count of 100. The example
spectrum shown in Figure 13-2 is of Cs137 and a vertical full-scale count of 1000.
Figure 13-2
To expand the vertical scale, from Figure 13-2, select the Zoom button on the menu bar and,
from the pop-up menu, select Vertical Zoom. This sequence can be repeated until the full-
scale count is at 200 as shown in Figure 13-3.

Table of Contents