EasyManua.ls Logo

Capintec CAPTUS 4000e - Making Measurements

Capintec CAPTUS 4000e
370 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CAPINTEC, INC. CAPTUS
®
4000e
9-16 ABSORBED DOSE ALERT August 15
Figure 9-16
1. Select the Delete button. Figure 9-17 will appear.
Figure 9-17
2. Select the Yes button or press the Y key to confirm the deletion or select the No button or
press the N key to cancel.
3. To delete all the patients at once, simply select the Delete All button. You will be
prompted to confirm the deletion. Figure 9-18 will appear.
Figure 9-18
4. select the Yes button or press the Y key to confirm the deletion. Otherwise, select the No
button or press the N key.
MAKING MEASUREMENTS
CAUTION: Prior to performing any measurements, it is recommended to carry out
the Auto Calibration procedure (Calibration, Linearity Correction and
Constancy Test only) described in CHAPTER 5: QUALITY
ASSURANCE, SECTION: AUTO CALIBRATION.

Table of Contents