EasyManua.ls Logo

Capintec CAPTUS 4000e - Making Measurements

Capintec CAPTUS 4000e
370 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CAPINTEC, INC. CAPTUS
®
4000e
15-12 CUSTOM PROTOCOLS December 15
MAKING MEASUREMENTS
CAUTION: Before making any measurements, calibrate the instrument for more
accurate results. Calibration is recommended every day. Refer to
CHAPTER 5: QUALITY ASSURANCE, for the procedure.
1. To select a patient for the test, click on the desired record in the Patient Directory list
box Figure 15-1 Custom Protocols Main Screen.
2. To begin the measurements, select the Count Selected Patients button or press
ALT+C. The screen appears as shown in Figure 15-15.
Note: If calibration has not been performed that day, you will be prompted with the
screen as in Figure 15-13. If calibration is performed, after the calibration is
complete, the program will return to the screen shown in Figure 15-14.
Figure 15-13
Figure 15-14

Table of Contents