CAPINTEC, INC. CAPTUS
®
4000e
November 16 GENERAL SETUP 4-17
Before performing any of the Quality Assurance tests, allow the system to warm up at least
30 minutes.
1. Select the Go button beside “Enter Check Source and Chi-Square Setup”. The Quality
Assurance Setup window will appear.
2. Enter the required fields. For details on the screen, see the pages listed in Reference
Pages below.
3. Upon exiting the Quality Assurance Setup window, you will be automatically returned to
the Startup Tips window.
4. Continue with the next test in the list (e.g., Auto Calibration).
5. Repeat steps 1-4 for each remaining test (Measure Efficiency, Constancy Test, and Chi-
Square Test).
6. If you do not wish to perform the Measurement Setups at this time, select the Exit button
or press Alt+E to close the Startup Tips window. Otherwise, continue to the
Measurements Setups section.
Note: If you do not want the Startup Tips window to appear the next time the CAPTUS
®
4000e software is started, click in the check box “Do Not Show This Screen
Again” at the bottom of the window after all setups have been completed.
Reference Pages
For further information on the tests, refer to the following pages in the manual:
Setup QA Parameters ....................................... 5-2
Perform Auto Calibration ................................... 5-5
Measure Efficiency of Check Source ................. 6-5
Perform Constancy Test .................................... 5-11
Perform Chi-Square Test ................................... 5-15
Measurement Setups
Note: It is only necessary to setup the tests that you are going to perform.
1. Select the Go button beside “Setup Thyroid Uptake Protocols”. The Setup Thyroid
Uptake Protocols screen will appear.
2. Enter the required fields. For details on the screen, see the pages listed in Reference
Pages below.
3. Upon exiting the Setup Thyroid Uptake Protocols screen, you will be automatically
returned to the Startup Tips window.
4. Continue with the next setup in the list (e.g., Wipe Test Defaults).