EasyManua.ls Logo

Capintec CAPTUS 4000e - Zero Offset; Threshold

Capintec CAPTUS 4000e
370 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CAPINTEC, INC. CAPTUS
®
4000e
13-16 MCA August 15
Figure 13-17
Gain is adjusted by typing the desired gain into in one or both of the Gain text boxes (Gain
range is 1 to 256).
In order to activate the newly entered gain, select the Done button or press Alt+D to return to
Figure 13-16. The System Gain number will update. This will set the gain until the MCA
module is exited.
Select the Cancel button or press Alt+C to discard the changes and return to Figure 13-16.
Zero Offset
The Zero Offset sets the channel location of zero keV. It is defined as percentage full-scale
with a range dependent on the CAP-MCA (Zero Offset range is -20.00 to +20.00%).
Use the scroll bar below the % Zero Offset indicator to adjust the value. Either click the right
(+) and left (-) arrows, or drag the scroll bar.
Threshold
The Threshold is set to eliminate any extraneous electronic noise while allowing counts of
the lowest possible energy to be detected. It is equivalent to a Lower Level Discriminator
(LLD). It is defined as percentage full-scale with a range dependent on the CAP-MCA
(Threshold range is 0.0 to 25.0%).
Use the scroll bar below the % Threshold indicator to adjust the value. Either click the right
(+) and left (-) arrows, or drag the scroll bar.
Note: The CAPTUS
®
4000e does not have a setting for ULD. The ULD is defined as
displayed spectrum full-scale (i.e., channel 1024, or 2MeV).

Table of Contents