10-10 LAB TESTS November 16
5. Select the Yes button or press the Y key to confirm the deletion. Otherwise, select the No
button or press the N key.
Making Measurements
CAUTION: Calibration is recommended before making any measurements. Refer to
CHAPTER 5: QUALITY ASSURANCE TESTS; SECTION: AUTO
• Before making any measurements on a patient, verify that all required patient information
has been entered.
• Make sure the default settings on the Edit Patient screen and the Setup screen are
appropriate. The values on Edit Patient screen override those on the Setup screen.
1. To select a patient for the test, click the patient in the Patient Directory box on Figure
10-2 Schilling Test Main Screen. The Patient Status box will show the current status of
the patient. If this box reads, “All Data has been entered”, the Count Selected Patients
button at the bottom of the screen is enabled and the test can be started. If data is
partially entered, the Count Selected Patients button will be disabled (grayed-out).
Note: If the Patient Status box shows “Data Partially Entered”, no measurements can
be made until all the required information is entered. Refer to Editing Patient
Information on page 10-7.
2. To begin the measurements, select the Count Selected Patients button or press Alt+C.
The screen appears as shown in Figure 10-12.
Note: If calibration has not been performed, you will be prompted with the screen as
shown in Figure 10-11. If calibration is performed, after the calibration is
complete, the program will return to the Schilling Test measurement screen
shown in Figure 10-12.
Figure 10-11