CAPINTEC, INC. CAPTUS
®
4000e
July 16 SAFETY 1-3
Located by each line cord on the Isolation Transformer is a label (Figure 1-2) indicating
which device each cord is to be connected to and the power availability for that device.
Figure 1-2
CAUTIONS AND NOTES
CAUTION: Changing the date or time while a timed test is in progress (for example,
a 24 hour Thyroid Uptake Test) might affect the results or introduce
unexpected error. Care should be taken to evaluate any potential impact
for ongoing tests before changing the date or time. Verify that in the Date
and Time Properties in Windows Control Panel that the “Automatically
adjust clock for daylight saving changes” checkbox is not checked.
CAUTION: Only qualified/trained personnel should operate or service this unit.
CAUTION: If the equipment is used in a manner not specified in this manual, the
protection provided by the equipment may be impaired.
CAUTION: High voltage is present inside the Detectors (up to 1000 Volts). Due to
the presence of this high voltage, opening the covers with the system
plugged in may be hazardous. Refer all servicing to qualified personnel.
CAUTION: The assembled CAPTUS
®
4000e system is very heavy (up to a
maximum of 740lb [336kg] fully loaded) Use caution when moving.
Reference CHAPTER 4: GENERAL SETUP, Section: Moving the
System for moving instructions.
CAUTION: There are no internal adjustments inside the detector, computer or
printer that should be performed by the user within the conditions of the
warranty, except for changing the Isolation Transformer fuse. Due to the
presence of high voltages, opening the cover of the computer with the
system plugged in may be hazardous. Refer all servicing to qualified