CAPINTEC, INC. CAPTUS
®
4000e
1-4 SAFETY July 16
CAUTION: Never use the Well Detector without the well liner in place. Liners are
inexpensive and easy to replace. A contaminated detector is a very
costly mistake. If unit becomes contaminated remove liner and clean unit
as stated in CHAPTER 16: CLEANING AND MAINTENANCE before
CAUTION: Care must be exercised when moving the system or when maintenance
is performed. If the system is inclined at more than a 15-degree angle, it
CAUTION: Verify that the collimator is correctly installed as stated in the
UNPACKING AND ASSEMBLY INSTRUCTIONS. Failure to follow these
instructions could result in injury or product malfunction.
CAUTION: This equipment generates radio frequency energy and is susceptible to
radio frequency energy (such as from other equipment and portable and
mobile RF communications equipment). The equipment needs to be
installed and used accordingly with these instructions; otherwise the
equipment may not function properly or may cause harmful
Electromagnetic Interference (EMI) to nearby devices. There is,
however, no guarantee that interference will not occur in a particular
CAUTION: If any Computer, Printer or cables other than the model supplied by
Capintec is used, the safety of the unit may be compromised or other
devices located in the same general area as the CAPTUS
®
4000e may
become susceptible to EMI.
CAUTION: If any detectors other than those supplied by Capintec are used, the
safety of the unit may be compromised or other devices located in the
same general area as the CAPTUS
®
4000e may become susceptible to
CAUTION: If any components are used other than those supplied by Capintec, the
power and signal connections must be the same as those for the original
manufacturer’s configuration or the safety of the unit may be
CAUTION: The equipment should not be used adjacent to or stacked with other
equipment and that if adjacent or stacked use is necessary, the
equipment should be observed to verify normal operation in the
configuration in which it will be used.
CAUTION: Do not disconnect or connect detectors from the computer, while the
system is on. Shut down the computer first.