July 16 SAFETY 1-5
CAUTION: Only the included USB devices should be connected to the USB ports on
the Computer. If any USB devices other than the models supplied by
Capintec are used, the safety of the unit may be compromised or other
devices located in the same general area as the CAPTUS
®
4000e may
become susceptible to EMI. The maximum Input/Output voltages are
5Vdc.
CAUTION: Do not attach a monitor to the VGA port on the rear of the Computer. If a
Monitor is attached to this port, the safety of the unit may be
compromised or other devices located in the same general area as the
CAPTUS
®
4000e may become susceptible to EMI.
CAUTION: Do not connect the CAPTUS
®
4000e to a network. If the system is
attached to a network, the safety of the unit may be compromised or
other devices located in the same general area as the CAPTUS
®
4000e
may become susceptible to EMI.
CAUTION: The unit contains lead. The unit should be disposed of in accordance
with local and national regulations.
CAUTION: The unit contains a lithium battery. This should be disposed of in
accordance with local and national regulations.
CAUTION: The user should always verify the validity of any measurement or test
result in order to minimize measurement errors.
CAUTION: To remove main power from the system, the Isolation Transformer power
cable must be unplugged from the medical grade wall socket.
WARNING: The CAPTUS
®
4000e must be connected a supply mains with protective
earth.
Note: It is recommended that periodic (every five years) re-calibration of the unit be
performed only by Capintec’s Authorized Service Center (reference CHAPTER 16:
CLEANING AND MAINTENANCE) to guarantee that the instrument's high reliability is
maintained).