EasyManua.ls Logo

Capintec CAPTUS 4000e - Counting Selected Patient

Capintec CAPTUS 4000e
370 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CAPINTEC, INC. CAPTUS
®
4000e
7-20 THYROID UPTAKE November 16
Counting Selected Patient
1. Select the Count Patient Dose button from Figure 7-1 Thyroid Uptake Main Screen. The
screen will appear as shown in Figure 7-16.
Figure 7-16
Note: If calibration has not been performed for the current day, you will be prompted
with the message window shown in Figure 7-17. If calibration is performed now,
the program will return to the Thyroid Uptake measurement screen shown in
Figure 7-16 after calibration is complete.
Figure 7-17
If Pre-Dose Measurement is set to Yesin the protocol, the message window will
appear as shown in Figure 7-18. To perform Pre-Dose count, select the Yes button or
press the Y key. The screen will appear as shown in Figure 7-16.

Table of Contents