EasyManua.ls Logo

Hioki 8835-01 - Setting for Measuring Voltage

Hioki 8835-01
406 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
127
────────────────────────────────────────────────────
7.3 Making the Settings of the 8936 ANALOG UNIT
────────────────────────────────────────────────────
7
.3.8 Configuring Baseline Offset
Adjusts voltage value to 0 V. Baseline offset is not available when range is set
above +,-10 DIV.
Baseline offset is available for measuring voltage and amperage. Not available
for any other measuring.
Configuration is not reset to initial settings when power is turned OFF.
New entry input resets settings.
Method Screen: CHANNEL 1, DISPLAY
1. Call up the CHANNEL 1 or DISPLAY screen.
2. Move the flashing cursor to the zero position
item, and select
zero setting.
: Opens baseline offset and zero adjustment
setting screen.
3. Use the function keys to make the selection.
: Applies baseline offset to all channels.
: Applies baseline offset to selected channel
only.
: Cancels baseline offset to selected channel.
: Disengages from zero adjustment setting
screen.
: Executes zero adjustment. Cancels baseline
offset to all channels.

Table of Contents

Related product manuals