.7.5 Setting for Measuring Voltage
NOTE
The settings of the waveform display color, waveform display graph type,
measurement range, input coupling, magnification/compression ratio along the
voltage axis, zero adjustment, baseline offset and low-pass filter are the same
as in the 8936 ANALOG UNIT.
For the settings, see Section 7.3.
Setting the Measurement Mode
Method Screen: CHANNEL 1
1. Call up the CHANNEL 1 screen.
2. Move the flashing cursor to the unit item.
3. Use the
JOG control or the function keys to
select Voltage.
: Move the cursor up in the selection window.
: Move the cursor down in the selection
window.
Setting the Pull-Up Resistance
See Section 7.7.2.
Probe Check
・When using the 9322 DIFFERENTIAL PROBE with the 8940 F/V UNIT,
the probe check confirms the 9322 connection and sets the voltage axis scale
multiplier to 1000 times. For frequency, duty ratio and count modes, the
threshold value is multiplied times 100.
・When several 9332s are connected to the 8940 F/V UNIT, the probe check
need be performed only once for all probes.
Method Screen: CHANNEL 1, DISPLAY
1. Move the flashing cursor to the unit item.
2. Use the
JOG control or the function keys to
select voltage. The probe check may be
performed from the frequency, count or duty
modes.
3. Connect the 9322 DIFFERENTIAL PROBE to
be used and select probe check from the
displayed function keys.
: Probe check is performed.
For more details, refer to instruction manual of the 9322 DIFFERENTIAL PROBE.