15.4 Input Cables
────────────────────────────────────────────────────
1
5.4.3 9018-10/9132-10 CLAMP-ON PROBE
DANGE
R
Always connect the clamp-on probe to the secondary side of a breaker.
On the secondary side of a breaker, even if the lines are shorted the
breaker can trip and prevent an accident. On the primary side, however
,
the current capacity may be large, and in the event of a short-circuit
there may be a serious accident.
Before using the 9018-10/9132-10 CLAMP-ON PROBE, be sure to
carefully read its instruction manual and familiarize yourself with the
operation principles of this product.
9132-1
0
9018-1
0
NOTE
The HIOKI 9018-10/9132-10 CLAMP-ON PROBE can be used to perform
current measurement. This probe has a voltage output. The 8835-01 is
designed for voltage input. It cannot be used to perform current measurements
by itself.
Connections
Connect the BNC connector of the 9018-10/9132-10 CLAMP-ON PROBE to the
analog input of the 8835-01. Refer to "2.4 Probe Connection."
The 8835-01 will indicate measurement results as voltage.
Using the scaling function, units can be converted to "A" for display. See
Section 9.4.
When the clamp-on probe is used for measurement, the measurement
precision will be affected both by the 8835-01 precision and clamp-on probe
precision ratings. The same is true for cases where other clamps are used.
When using the 9018/9132 CLAMP-ON PROBE, always use the 9199
CONVERSION ADAPTER.