275
────────────────────────────────────────────────────
12.6 Using the Evaluation Outputs (GO), (NG)
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
CAUTIO
N
To prevent damage to the unit, take care never to exceed the voltage rating of
the GO and NG outputs.
Output
Evaluation result
GO NG
GO
Low level High level
NG
High level Low level
4.7 kΩ
5V
10 kΩ
4.7 kΩ
Hig
h
4.0 to 5.0 V
Low
0 to 0.5 V
min. 70 m
s
Evaluation in progres
s
GO
―――
,NG
―――
1
2.6 Using the Evaluation Outputs (GO), (NG)
When waveform evaluation or waveform parameter evaluation is used, a
signal is output from these connectors when the result is GO (pass) or NG
(fail).
Output signal
Signal type open-collector signal, active Low
Output voltage High level: 4.0 to 5.0 V
range Low level: 0 to 0.5 V
Maximum input voltage -20 to +30 V, max. 100 mA, max. 200 mW
Evaluation output interval (min. 70 ms)
The evaluation outputs are shown in the following table. Between these states,
there is an interval during which the next data are read and waveform data
are created. The duration of this interval is inversely proportional to the time
axis and proportional to the recording length.