339
────────────────────────────────────────────────────
15.4 Input Cables
────────────────────────────────────────────────────
When using grabber clips, the 9322’s maximum rated voltage to earth is
1500 V AC/DC; when using alligator clips, it is 1000 V AC/DC. To avoid
electrical shock and possible damage to the unit, never apply voltages
greater than these limits between the input channel terminals and
chassis, or across the inputs of two 9322s.
Maximum input voltage is 1000 VAC/2000 VDC. Do not measure voltage
in excess of these limitations, as doing so may damage the unit or
cause an accident that might result in injury or death.
Grabber clip
9322 is a differential probe that connects to input of 8835-01 MEMORY
HiCORDER input unit. For more details, refer to its instruction manual.