7.7 Making the Settings of the 8940 F/V UNIT
────────────────────────────────────────────────────
Dead time50 μs
can not be measure
d
Displays the measurement mod
e
7
.7.1 Settings
7
.7 Making the Settings of the 8940 F/V UNIT
This section explains the settings of the 8940 F/V UNIT.
The 8940 F/V UNIT measures frequency, count, pulse duty ratio, voltage and
current at each channel (one measurement type is available for each channel).
Frequency: Frequency is obtained from input pulses derived from the
measured waveform.
Count: This is a count of the number of input pulses.
Pulse Duty Ratio: This is the duty ratio of the measured waveform.
Voltage: Measures the voltage waveform
Current: Measures the current waveform
Notes on measuring the pulses duty ratio
Upon measurement of pulses that rise during dead time, the duty ratio is
determined from pulses that follow the dead time; the duty ratio of dead time
pulses is not measured.
The settings of the waveform display color, waveform display graph type,
magnification/compression ratio along the voltage axis, zero adjustment,
baseline offset and low-pass filter are the same as in the 8936 ANALOG
UNIT.
For the settings, see Section 7.3.