176
────────────────────────────────────────────────────
8.5 Setting Trigger Source AND/OR Linking
────────────────────────────────────────────────────
NOTE
Channe
.5 Setting Trigger Source AND/OR Linking
The analog trigger, logic trigger, and timer trigger can be linked with the
AND/OR logical operators.
OR: Triggering occurs when conditions for one trigger are met.
AND: Triggering occurs when conditions for all triggers are met.
If the trigger source is set to AND, and the trigger source setting conditions
have already been met when the
START key is pressed, triggering does not
occur. When the conditions are not met once but met subsequently, triggering
occurs.
Method Screen: STATUS 2
1. Press the STATUS key to call up the
STATUS 2 screen.
2. Move the flashing cursor to the trigger source
item, as shown in the figure on the left.
3. Use the function keys to select the AND or
OR.
: Link trigger sources with logical operator
OR.
: Link trigger sources with logical operator
AND.
Example The figures below show the difference between the effect of AND/OR linking.